User’s and Service Guide U3047-90001
59
U3047AM12
Service Information
Troubleshooting the Test Set
If the U3047AM12 is not operating properly, use the following procedures to isolate and repair the type of
failure encountered. It is recommended that a qualified service technician perform the following
procedures.
Refer to the Agilent PNA Series: Service & Support Home Page at:
http://na.tm.agilent.com/pna
for
further information.
To request service, please contact your local service center. In the US, call 800-829-4444. For a listing of
service centers worldwide, please visit us at www.agilent.com/find/service.
WARNING
No operator serviceable parts inside. Refer servicing to qualified personnel. To prevent
electrical shock do not remove covers.
Non-RF Failures
Typical malfunctions may be the rear panel fan, control lines or front panel indicator lights.
1. Verify that the front panel Power Switch is operational.
a. The rear panel fan and front panel Active LED should operate when the Standby Switch is in the On
position.
2. AC line voltage checks (remove AC power from the instrument).
a. Ensure the proper AC line voltage is present at the instrument line cord.
b. Remove the AC power cord from the instrument. Confirm the instrument AC line module fuse is
operational. See
.
c. Internal AC line fuses - Remove the bottom cover of the instrument. Near the rear panel are two
fuse holders, remove the fuses and verify that they are operational. Refer to
.
WARNING
The opening of covers or removal of parts is likely to expose the user to dangerous
voltages. Disconnect the instrument from all voltage sources before being opened.
3. Internal DC Power - During this check you will apply AC power to the instrument with the top cover
removed. Connect the AC power cord to the test set.
a. Set the front panel switch to the Standby position. The fans and indicator lights should be Off.
b. Set the standby switch to the On position. The rear panel and internal power supply fans should be
operational. The Active LED will be On when test set is connected to an active network analyzer and
is addressed by the network analyzer
c. Verify that the DC LEDs indicators are On.
Note: The LEDs only indicate the presence of a DC voltage, not the DC voltage level. Use the
indicated wire locations for voltage measurements, TP1 is your common. Refer to
.
d. If there is no DC power, suspect the power supply module. The actual power supply terminal
connection should also be confirmed with a DVM by removing the bottom cover and using the
power supply label to identify the terminal voltage values.
e. If all of the DC Input LEDs are On and the fan is not working, replace fan.