8
User’s and Service Guide U3047-90001
U3047AM12
General Specifications
Maximum Power Levels
CAUTION
It is recommended that you do not operate components near damage levels
(+28 dBm). The power levels must be 3 dB below maximum level to ensure no damage.
NOTE
Refer to your analyzer’s specifications to determine the maximum input power levels for the
access and test ports, or to optimize the power levels in the receivers.
NOTE
Damage and maximum levels are not necessarily the optimum level.
Table 4
Maximum Power Levels
U3047AM12 Test Port RF Power Levels:
PORT 5-16
+25 dBm
U3047AM12 Access Ports:
SOURCE OUT
+15 dBm
CPLR ARM
+15 dBm
CPLR THRU
+15 dBm
RCVR OUT
+15 dBm