User’s and Service Guide U3047-90001
33
U3047AM12
Controlling the Test Set
Using Interface Control Mode
An Instrument Preset will reset all of the fields to their default settings.
NOTE
If an error is encountered when sending Interface Control data, an error message is
displayed on the network analyzer screen and the Channel Trigger State is set to Hold. You
must fix the condition that caused the error, then change the Channel Trigger State to its
original setting
Figure 31
Interface Control
Enable Interface Control: (1)
Enables and disables ALL Interface Control communication. When cleared (default setting) Interface
Control is disabled and NO data is sent. To send data, the individual interfaces must also be enabled.
Test Set I/O Control (addr.data): (2)
Provides control of the test set I/O Interface on the rear panel of the network analyzer. Used to control your
test set.
Only positive integers are allowed to select switch positions or states of DUT control interface lines. Refer
to
“Address and Data Values” on page 38
.
Address and data integers must be separated by a period. Each set of entries should be separated by a new
line, or carriage return. The front panel Enter key can be used to insert a new line into the field. The
quantity of test set I/O entries that can be entered is limited by the available memory of the network
analyzer.
Address and Data example:
addr.data
(3)
0.0
16.1
32.2
(2)
(3)
(5)
(7)
(8)
(8)
(6)
(1)
(4)