24
User’s and Service Guide U3047-90001
U3047AM12
Controlling the Test Set
RF Path Configuration
NOTE
If the analyzer has Option 029 or H29 (noise figure capability) verify that the path
configuration is set to Default. If the network analyzer is not in Default configuration, ports
5 and 9 will not operate correctly. Select
Trace/Chan > Channel
>
Hardware Setup > Path
Cong
... and in the drop-down menu select
Default
>
OK
. Save this configuration as a User
Preset by selecting
Save
>
User Preset
>
Save current state as User Preset
. Do not use
the factory Preset (User Preset Off), the analyzer will return to Option 029 path
configuration.
Figure 21
RF Path Configuration