User’s and Service Guide U3047-90001
7
U3047AM12
General Specifications
Environmental Requirements
Refer to the PNA series standard documentation for environmental requirements.
Environmental Tests
The test set complies with all applicable safety and regulatory requirements for the intended location of
use.
•
Temperature 0 - 50 °C
•
Pressure Altitude
3,000 meters (~10,000 feet)
•
The instrument can safely operate in a relative humidity of 80% for temperatures to
31 degrees C, decreasing linearly to 50% relative humidity at 40 degrees C.
Equipment Heating and Cooling
If necessary, install air conditioning and heating to maintain the ambient temperature within the
appropriate range.
CAUTION
Ventilation Requirements: When installing the instrument in a cabinet, the convection into
and out of the instrument must not be restricted. The ambient temperature (outside the
cabinet) must be less than the maximum operating temperature of the instrument by 4 °C
for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is
greater than 800 watts, forced convection must be used.
Required Conditions for Accuracy Enhanced Measurement
Accuracy–enhanced (error–corrected) measurements require the ambient temperature of the network
analyzer and test set to be maintained within ± 1 °C of the ambient temperature at calibration.
Dimensions and Space Requirements
Standard installation of the test set and network analyzer includes configuration and installation on a
customer provided lab bench or table top of adequate size and strength. For weight, dimensions and space
requirements, refer to the network analyzer documentation that is used to configure the test set.
CAUTION
The network analyzer is heavy. It is recommended that two individuals, or a mechanical lift
be used to lift or transport the instrument.
Table 3
Instrument Dimensions
Model
Weight
Height
Width
Depth
U3047AM12
10 kg (22 lb)
22.15 cm
(8.7 in)
42.9 cm
(16.9 in)
52.4 cm
(20.6 in)