64
User’s and Service Guide U3047-90001
U3047AM12
Service Information
Cable Connections
The RF flex cables will be connect to the designated test set front panel ports and an uncorrected response
trace should resemble the figures indicated in
Source Signal Path Insertion Loss Test
Connect the RF Flex cables to the Test Port and Source IN port indicated in
, the expected results
should be similar to
through
.
Table 16
RF Signal Path Insertion Loss (S21)
RF Path Description
Signal
Insertion Loss (typical)
Source IN to Ports 5-8
Source
Source IN to Ports 9-12
Source
Source IN to Ports 13-16
Source
RCVR OUT to Ports 5-8
Receiver
RCVR OUT to Ports 9-12
Receiver
RCVR OUT to Ports 13-16
Receiver
Source IN to CPLR THRU
Source
RCVR OUT to CPLR ARM
Receiver
Table 17
Source Signal Path Test Instructions
Path #
RF Path
Description
Control Mode
<Address>.<Data>
Path Components
Insertion Loss (typical)
1
Source IN to Port 5
1
1. Use the Source IN port associated with this group of test ports.
0.1
P5 CPLR, S10 & 11
2
Source IN to Ports9
0.2
P9 CPLR, S10, 11 & 12
3
Source IN to Port 13
0.8
P13 CPLR, S10, 11, 12 & 16
4
Source IN to Port 6
16.1
P6 CPLR, S20 & 21
5
Source IN to Port10
16.2
P10 CPLR, S20, 21 & 22
6
Source IN to Port 14
16.8
P14 CPLR, S20, 21, 22 & 26
7
Source IN to Port 7
32.1
P7 CPLR, S30 & 31
8
Source IN to Port 11
32.2
P11 CPLR, S30, 31 & 32
9
Source IN to Port 15
32.8
P15 CPLR, S30, 31, 32 & 36
10
Source IN to Port 8
64.1
P8 CPLR, S40 & 41
11
Source IN to Port 12
64.2
P12 CPLR, S40, 41 & 42
12
Source IN to Port 16
64.8
P16 CPLR, S40, 41, 42 & 46
Содержание U3047AM12
Страница 4: ...iv...
Страница 7: ...User s and Service Guide U3047 90001 1 U3047AM12...