Performance Tests
1
Verification Manual
91
52
step 13
‘No Trouble’ Max Sens
Pass/Fail
52
step 15
Power Meter check
–10.5
–9.5
49
step 5
Laser LED ON
Pass/Fail
49
step 6
Note Attenuator
Setting
N/A
N/A
49
step 9
‘No Trouble’ Min Sens
Pass/Fail
50
step 10
Additional Attenuation
Typ > 1 dB
N/A
50
step 11
–30 dBm PWR meas
–31 dBm
–29 dBm
50
step 11
OC-3 /STM-1
Frequency
155.51999 MHz
155.52001 MHz
51
step 12
–10 dBm PWR meas
–11 dBm
–9 dBm
52
step 13
‘No Trouble’ Max Sens
Pass/Fail
52
step 15
Power Meter check
–10.5
–9.5
Optical Receive Port (622 Mb/s)
52
step 17
Laser LED ON
Pass/Fail
52
step 18
Note Attenuator
Setting
N/A
N/A
53
step 21
‘No Trouble’ Min Sens
Pass/Fail
53
step 22
Additional Attenuation
Typ > 1 dB
N/A
53
step 23
–28 dBm PWR meas
–29 dBm
–27 dBm
53
step 23
OC-12/STM-4
Frequency
622.07999 MHz
622.08001 MHz
53
step 24
–8 dBm PWR meas
–9 dBm
–7 dBm
53
step 25
‘No Trouble’ Max Sens
Pass/Fail
53
step 27
Power Meter check
–8.5
–7.5
Optical Receive Port (2.5 Gb/s)
54
step 3
Laser LED ON
Pass/Fail
Page No.
Test Description
Min
Result
Actual
Max