28
Verification Manual
1
Performance Tests
14
Ensure the PDH/DSn test set transmitter is set as follows.
SIGNAL to 2 Mb/s rate
CLOCK to EXTERNAL 10 MHz REF
TERMINATION to 75 ohm UNBAL
LINE CODE to HDB3
PAYLOAD TYPE to UNFRAMED
PATTERN to 2^23 PRBS
15
Check that the test set CLOCK LOSS Alarm LED is not lit.
16
Adjust the frequency counter trigger level to obtain a stable
reading and check that the frequency counter reads between:
10MHz Reference to External 10M Clock Input
PDH/DSn Test Set
2 Mb/s DATA
2 MHz Clock O/P
step 18
Matching pad
and/or termination
53181A Counter
Counter 10 MHz
Reference OUTPUT
OmniBER OTN
UNBAL
BAL