![Agilent Technologies OmniBER OTN J7230B Скачать руководство пользователя страница 73](http://html1.mh-extra.com/html/agilent-technologies/omniber-otn-j7230b/omniber-otn-j7230b_performance-verification-manual_2868192073.webp)
70
Verification Manual
1
Performance Tests
.
3
Load the
2Mb_ITU_75.msk
test mask and check the Positive
and Negative pulses are within the mask (see examples).
Adapter
54810A Oscilloscope
OmniBER OTN
CH1
BNC to BNC cable
E2622A
75/50
Ω
2/8/34/140 Mb/s & DS-3 UnBal OUT
N O T E
To test Negative pulse set the Scope
Mask Test Set-up
panel to
Invert
Mask
Verification.book Page 70 Monday, January 19, 2004 1:40 PM
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com