Performance Tests
1
Verification Manual
57
OC-192/STM-64 Rate (J7230B and J7232A Opt 108 Only)
1
Press
<Menu>
, choose
System > Stored Settings
then press
<Select>
. Select
Default
then
Recall
.
2
Connect the test setup as shown
below.
3
Set the optical attenuator to:
ATTEN 15 dB
WAVELENGTH 1550 nm
CAL=0
ENB ON
Set the optical power meter to
WAVELENGTH 1550 nm
Connect the optical attenuator between the power meter and
the OmniBER OTN’s
Optical Out
port (ensure that all
connections are tight and that the cable has no twists).
9.95 Gb/s Optical Output
8156A Optical Attenuator
8163/81633A Optical Power Meter
OmniBER OTN
Optical Out
Optical In
step 6