56
Verification Manual
1
Performance Tests
12
Disable the optical attenuator then disconnect the Optical
signal from the OmniBER OTN
Optical IN
port and connect
to the power meter. Re-enable the optical attenuator.
13
Check the power meter reading is between:
–8.5 dBm and –7.5 dBm
Verification.book Page 56 Monday, January 19, 2004 1:40 PM