40
Verification Manual
1
Performance Tests
9.95 Gb/s and 10.71 Gb/s Optical Output Checks (J7230B and J7232A
Opt 108 Only)
OC-192/STM-64 Power Level and Eye Mask Tests
1
Press
<Menu>
, choose
System > Stored Settings
then press
<Select>
. Select
Default
then
Recall
.
2
Press
<Menu>
, choose
Tx/Rx > Transmitter Settings >
Physical
then press
<Select>
. Set the
Signal Mode
to
SONET
and
Signal Rate
to
OC192
. Ensure the
Laser
is set to
OFF
.
3
Connect the
9.95/10.71 Gb/s Optical Out
port to the
oscilloscope using an Optical Patch cord. Also connect the
connect the
TX
Eye Clock
(9.95/10.71 Gb/s) to the
oscilloscope Trigger input as shown.