Performance Tests
1
Verification Manual
27
7
Confirm that the OmniBER OTN is locked to the external
signal by disconnecting/re-connecting the BNC cable at the
instrument CLOCK IN port
(External 10M Clock as Reference)
8
Now set the instrument to
External Clock, 10M Ref.
9
Disconnect the 10M Reference signal from the PDH/DSn test
set and connect it directly to the OmniBER OTN as shown in
the diagram above.
10
Repeat
step 5
through
step 7
(External 2M Unbalanced Data as Reference)
11
Re-connect as in the diagram below.
12
Change the OmniBER OTN
External Clock
,
Format
setting
to
2 Mb/s Data
.
13
Change the 2 MHz CLOCK OUT connection at the PDH/DSn
test set to connect the transmit 2 Mb/s DATA out BNC to the
OmniBER OTN CLOCK IN BNC port.