Commonly Used Procedures 6
Agilent Nano Indenter G200 User’s Guide
6-34
Soft polymers are the primary application for flat-ended cylidrical
indenters.
Users may add new tip definitions or delete existing ones using the
Add
and
Remove
buttons as shown in
.
Figure 6-48
Use Add and Remove buttons in Configure Tip Parameters
dialog box
As shown in
, users may choose to:
•
Calibrate Tip and Frame
(most common)
•
Calibrate Frame (less common)
N O T E
Available tip definitions are stored in NanoSuite. The actual tip
definition for a particular sample is stored in the .mss file. So even if a
tip definition is deleted in NanoSuite, it is not deleted from the sample
file. If a sample file is opened and the tip definition doesn’t exist in
NanoSuite, the software will prompt you to save the definition.
Содержание Nano Indenter G200
Страница 1: ...Agilent Technologies Agilent Technologies Nano Indenter G200 User s Guide...
Страница 14: ...Agilent Nano Indenter G200 User s Guide 9 Express Test Tip Calibration H 12 Express Test Varied Force H 12...
Страница 96: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 4 Export Menu Mode Menu System Menu Help Menu...
Страница 109: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 17 Figure 4 18Load Review page configuration dialog box...
Страница 117: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 25 Figure 4 27Real Time Graph Setup dialog box...
Страница 121: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 29 Figure 4 31User Configuration dialog box...
Страница 136: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 44 Exit Use this icon to exit the software...
Страница 263: ...Commonly Used Procedures 6 Agilent Nano Indenter G200 User s Guide 6 44 Figure 6 62Flowchart 1 Figure 6 63Flowchart 2...
Страница 478: ...Agilent Technologies Agilent Nano Indenter G200 User s Guide Part Number G2A 13192 1 Rev C Agilent Technologies Inc 2012...