Quick Start Guide 2
Agilent Nano Indenter G200 User’s Guide
2-25
Figure 2-31
Thumbwheel to adjust height for rear left sample disk
6
Turn the thumbwheel from left to right to lower the sample disk as
far as it will go down but not lower than the surface of the sample
tray.
7
Turn the brass thumbscrew just enough to secure the disk in the hole
but not so tight that the disk cannot travel vertically in the sample
tray.
8
Right-click anywhere on the schematic map and select
Nano Video
Handset
from the resulting pop-up menu.
The image will probably be out of focus at this point.
N O T E
Each brass thumbscrew contains a spring-loaded ball plunger that
applies pressure to the disk. Do not fully tighten the thumbscrew at this
stage because the disk will need to be adjusted vertically during a
subsequent stage.
Содержание Nano Indenter G200
Страница 1: ...Agilent Technologies Agilent Technologies Nano Indenter G200 User s Guide...
Страница 14: ...Agilent Nano Indenter G200 User s Guide 9 Express Test Tip Calibration H 12 Express Test Varied Force H 12...
Страница 96: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 4 Export Menu Mode Menu System Menu Help Menu...
Страница 109: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 17 Figure 4 18Load Review page configuration dialog box...
Страница 117: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 25 Figure 4 27Real Time Graph Setup dialog box...
Страница 121: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 29 Figure 4 31User Configuration dialog box...
Страница 136: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 44 Exit Use this icon to exit the software...
Страница 263: ...Commonly Used Procedures 6 Agilent Nano Indenter G200 User s Guide 6 44 Figure 6 62Flowchart 1 Figure 6 63Flowchart 2...
Страница 478: ...Agilent Technologies Agilent Nano Indenter G200 User s Guide Part Number G2A 13192 1 Rev C Agilent Technologies Inc 2012...