Instrument Reference 3
Agilent Nano Indenter G200 User’s Guide
3-8
Simple contact points are added to the end of the torsion-bar cantilever
and the loading element, so that upon contact between the torsion bar
and the loading element a circuit is completed. Once the contact
between the torsion bar and the loading elements is recognized, load is
applied by applying torsion to the bar. Unloading is accomplished by
releasing the torque on the torsion bar. When the contact-recognition
circuit is broken, the cantilever is moved away from the loading
elements.
, "High Load" for more information.
Indenter
The most common geometry for the diamond tip used with the
indentation head is the Berkovich indenter. Shown in
Berkovich indenter is a three-sided pyramid with the same aspect ratio
as the Vickers indenter. Any diamond indenter geometry that is desired
can be used with the indenter head, provided that the diamond mount is
configured to fit into the indenter shaft.
Figure3- 10
Diamond indenter with Berkovich geometry
Available tip geometries include conical, Knoop, Vickers, and wedge.
Contact Agilent for more information about alternative diamond-tip
geometries. Also, see
Isolation Cabinet
The G200 isolation cabinet (
laminate exterior and a foam interior. The isolation cabinet is used to
retard environmental temperature changes and dampen acoustic
disturbances. The closed (and sealed) cabinet provides a stable air mass
in which sudden temperature changes are minimized. The foam lining
on the interior of the cabinet absorbs acoustic energy and decreases the
transmission of such excitation to the instrument. The fixed cable ports
Содержание Nano Indenter G200
Страница 1: ...Agilent Technologies Agilent Technologies Nano Indenter G200 User s Guide...
Страница 14: ...Agilent Nano Indenter G200 User s Guide 9 Express Test Tip Calibration H 12 Express Test Varied Force H 12...
Страница 96: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 4 Export Menu Mode Menu System Menu Help Menu...
Страница 109: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 17 Figure 4 18Load Review page configuration dialog box...
Страница 117: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 25 Figure 4 27Real Time Graph Setup dialog box...
Страница 121: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 29 Figure 4 31User Configuration dialog box...
Страница 136: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 44 Exit Use this icon to exit the software...
Страница 263: ...Commonly Used Procedures 6 Agilent Nano Indenter G200 User s Guide 6 44 Figure 6 62Flowchart 1 Figure 6 63Flowchart 2...
Страница 478: ...Agilent Technologies Agilent Nano Indenter G200 User s Guide Part Number G2A 13192 1 Rev C Agilent Technologies Inc 2012...