NanoSuite Explorer D
Agilent Nano Indenter G200 User’s Guide
D-48
Figure D-58
New description for Poisson’s Ratio
Sometimes you may wish to find where a particular input or formula is
used in the method. For example, how is the Poisson’s Ratio input used
in this method? There is an efficient way to do this:
1
Save
the method.
2
Highlight the input
Poissons Ratio
in the Settings pane.
3
Right-click the selection.
4
Select
Delete
from the resulting pop-up menu.
This produces an error in the
5
Double-click on the error message, which takes you directly to the
point where the method requires this input.
6
To see exactly how the input is used in this case, click the
Formula
tab.
7
Close
the current method by opening any other method.
8
When
asked if you want to save changes to the current method, click
No
.
Exercise 2
Create a channel that calculates strain rate, defined as displacement rate
divided by displacement. For some materials, the imposed strain rate
may affect the measured hardness, so it would be nice to keep track of
this parameter. Also, open another previously tested sample and import
it into the modified method.
1
Launch NanoSuite, logging in as a user with Creator priveleges.
2
Choose
File > Open Sample...
.
Содержание Nano Indenter G200
Страница 1: ...Agilent Technologies Agilent Technologies Nano Indenter G200 User s Guide...
Страница 14: ...Agilent Nano Indenter G200 User s Guide 9 Express Test Tip Calibration H 12 Express Test Varied Force H 12...
Страница 96: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 4 Export Menu Mode Menu System Menu Help Menu...
Страница 109: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 17 Figure 4 18Load Review page configuration dialog box...
Страница 117: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 25 Figure 4 27Real Time Graph Setup dialog box...
Страница 121: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 29 Figure 4 31User Configuration dialog box...
Страница 136: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 44 Exit Use this icon to exit the software...
Страница 263: ...Commonly Used Procedures 6 Agilent Nano Indenter G200 User s Guide 6 44 Figure 6 62Flowchart 1 Figure 6 63Flowchart 2...
Страница 478: ...Agilent Technologies Agilent Nano Indenter G200 User s Guide Part Number G2A 13192 1 Rev C Agilent Technologies Inc 2012...