Commonly Used Procedures 6
Agilent Nano Indenter G200 User’s Guide
6-51
Frame Stiffness Calibration: ISO 14577 Method
1
Make sure that the fused silica standard is in the center sample-disk
hole in the sample tray.
2
In NanoSuite, select
Open Method…
from the Method menu.
3
From the Filename drop-down box in the Open Method dialog box,
select
ISO Standards > XP ISO 14577 Standard Test Method
as
Figure 6-72
Selecting the ISO method
4
Click
OK
.
5
Select
Batch Mode
from the Mode menu.
6
Right-click in the right pane in NanoSuite and select
Load Batch
.
7
In the Retrieve Batch File dialog box, open
C:\Program Files\
Agilent NanoSuite\Methods\ISO Standards\Machine
Compliance Batch.bch
.
8
Click
Open
.
9
Click
Run Batch
on page 6-52) to initiate testing.
Содержание Nano Indenter G200
Страница 1: ...Agilent Technologies Agilent Technologies Nano Indenter G200 User s Guide...
Страница 14: ...Agilent Nano Indenter G200 User s Guide 9 Express Test Tip Calibration H 12 Express Test Varied Force H 12...
Страница 96: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 4 Export Menu Mode Menu System Menu Help Menu...
Страница 109: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 17 Figure 4 18Load Review page configuration dialog box...
Страница 117: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 25 Figure 4 27Real Time Graph Setup dialog box...
Страница 121: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 29 Figure 4 31User Configuration dialog box...
Страница 136: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 44 Exit Use this icon to exit the software...
Страница 263: ...Commonly Used Procedures 6 Agilent Nano Indenter G200 User s Guide 6 44 Figure 6 62Flowchart 1 Figure 6 63Flowchart 2...
Страница 478: ...Agilent Technologies Agilent Nano Indenter G200 User s Guide Part Number G2A 13192 1 Rev C Agilent Technologies Inc 2012...