42
N9310A User’s Guide
3
Using Functions
Generating an AM signal
The signal generator generates AM (amplitude modulated)
signals with the following basic settings:
Operation Example
Assume that you are going to generate an AM signal with:
• Carrier frequency of 1 GHz, amplitude of –10 dBm
• AM depth of 70 %
• AM rate at 15 kHz
• Internal AM source (default by
)
Refer to the following to configure your signal generator:
1
Pressing
presets the signal generator
2
Pressing
>
1
>
GHz
sets
the
CW
frequency
to
1
GHz
3
Pressing
> –10 > dBm
sets the
CW
amplitude
to
–10 dBm
4
Pressing
enters the AM submenu
5
Pressing
AM Depth > 70 > %
sets AM depth to 70 %
6
Pressing
AM Rate > 15 > kHz
sets AM rate to 15 kHz
7
Pressing
AM On
enables AM
8
Pressing
On
enables the AM signal output
For key reference, please refer to
.
Character
Range
Default
AM Depth
0.0 to 100.0 %
0.0
AM Source
Internal source (INT)
External source (EXT)
Combined INT and EXT
INT
AM Rate
20 Hz to 80 kHz (INT)
DC/20 Hz to 80 kHz (EXT)
1.0000 kHz, 0.1 Hz resolution
Preset
Preset
Frequency
Amplitude
AM
RF
On/Off
Содержание Keysight N9310A
Страница 34: ...28 N9310A User s Guide 2 Getting Started ...
Страница 58: ...52 N9310A User s Guide 3 Using Functions ...
Страница 94: ...88 N9310A User s Guide 5 Programming Fundamentals 5 Select Display a list ...
Страница 106: ...100 N9310A User s Guide 5 Programming Fundamentals ...
Страница 168: ...162 N9310A User s Guide 6 Subsystem Command Reference ...
Страница 188: ... Agilent Technologies Inc 2006 Printed in China June 2006 N9310 90001 N9310 90001 ...