72
N9310A User’s Guide
4
Key Reference
Pulse Period
Pressing this softkey modifies the pulse period of the
internal pulse signal. The active function area displays the
current pulse period value. The allowed value range is 200
μ
s to 2 s, with the minimum increment of 1
μ
s. The pulse
period applies only to the internal pulse signal. If an
external pulse source is selected, the
Pulse Period
softkey
is
disabled.
•
Default value:
200
μ
s
•
Key sequence:
>
Pulse Period
Pulse Width
Pressing this softkey modifies the pulse width for the
internal pulse signal. The active entry area displays the
current pulse width value. The allowed value range is 100
μ
s
to 1 s, with the minimum increment of 1
μ
s. If an external
pulse source is selected, the
Pulse Width
softkey is disabled.
•
Default value:
100
μ
s
•
Key sequence:
Pulse
>
Pulse Width
Pulse