36
N9310A User’s Guide
3
Using Functions
Generating an RF Sweep
Assume you are going to generate an RF sweep with the fol-
lowing settings:
• frequency range from 1 GHz to 2 GHz, at a level of 0
dBm
• nine sweep points and 50 ms dwell time at each point
Refer to the following procedures to configure and generate
an RF sweep:
1
Press
hardkey
2
Press
>
0
>
dBm
This step sets the RF sweep output level at 0 dBm.
3
Press
hardkey
This step opens the sweep submenus.
4
Press
Step Sweep
softkey
This step opens a submenu for step sweep configurations.
5
Press
RF Start > 1 > GHz
This step sets the start frequency of the sweep to 1 GHz.
6
Press
RF Stop > 2 > GHz
This step sets the stop frequency of the sweep to 2 GHz.
7
Press
#
Points > 9 > Enter
This step sets the number of sweep points to nine.
8
Press
More(1/2) > Step Dwell > 50 > ms
This step sets the dwell time at each point to 50 ms.
9
Press
hardkey to enable the RF output.
The annunciator changes from
to
.
Preset
Amplitude
Sweep
RF
On/Off
Содержание Keysight N9310A
Страница 34: ...28 N9310A User s Guide 2 Getting Started ...
Страница 58: ...52 N9310A User s Guide 3 Using Functions ...
Страница 94: ...88 N9310A User s Guide 5 Programming Fundamentals 5 Select Display a list ...
Страница 106: ...100 N9310A User s Guide 5 Programming Fundamentals ...
Страница 168: ...162 N9310A User s Guide 6 Subsystem Command Reference ...
Страница 188: ... Agilent Technologies Inc 2006 Printed in China June 2006 N9310 90001 N9310 90001 ...