Subsystem Command Reference
6
N9310A User’s Guide
107
LF Start
:FREQuency:LF:STARt <val> <unit>
:FREQuency:LF:STARt?
This command sets the start frequency in an LF sweep. The
start frequency should always be lower than the stop
frequency in an LF sweep.
The query returns the current start frequency.
LF Stop
:FREQuency:LF:STOP <val> <unit>
:FREQuency:LF:STOP?
This command sets the stop frequency in an LF sweep. The
stop frequency should always be set higher than the start
frequency.
The query returns the current stop frequency.
Range
20 Hz to 80 kHz
Valid unit
kHz, Hz
Minimum increment
0.1 Hz
*RST value
20.0 Hz
Returned data format
<val> <unit>
Front panel access
Sweep >
Step Sweep > LF Start
Range
20 Hz to 80 kHz
Valid unit
kHz, Hz
Minimum increment
0.1 Hz
*RST value
80.0000 kHz
Returned data format
<val> <unit>
Front panel access
Sweep > Step Sweep > LF Stop