Subsystem Command Reference
6
N9310A User’s Guide
119
Sweep Repeat
:SWEep:REPeat SINGle|CONTinuous
:SWEep:REPeat?
This command toggles the sweep repeat between single and
continuous.
The query returns the current sweep repeat state.
Sweep Trigger
:SWEep:STRG IMMediate|KEY|EXT
:SWEep:STRG?
This command sets the sweep trigger mode to one of the
following choices:
•
IMMediate
- Selecting this mode immediately triggers
signal generator
to sweep from the first point to the final
point in a sweep sequence automatically.
•
KEY
- Selecting this mode arms a sweep. The armed
sweep will be initiated by sending the
“
TRIGger:IMMediate
” command.
•
EXT
- Selecting this mode enables an external trigger
source to initiate an armed sweep.
The query returns the current sweep trigger mode.
Range
SINGle/CONTinuous
*RST value
CONTinuous
Returned data
SING, CONT
Front panel access
Sweep > Sweep Repeat
Range
IMMediate, KEY, EXT
*RST value
IMMediate
Returned data
IMM, KEY, EXT
Front panel access
Sweep > Sweep Trigger
Содержание Keysight N9310A
Страница 34: ...28 N9310A User s Guide 2 Getting Started ...
Страница 58: ...52 N9310A User s Guide 3 Using Functions ...
Страница 94: ...88 N9310A User s Guide 5 Programming Fundamentals 5 Select Display a list ...
Страница 106: ...100 N9310A User s Guide 5 Programming Fundamentals ...
Страница 168: ...162 N9310A User s Guide 6 Subsystem Command Reference ...
Страница 188: ... Agilent Technologies Inc 2006 Printed in China June 2006 N9310 90001 N9310 90001 ...