Using Functions
3
N9310A User’s Guide
39
Generating an LF Sweep
Assume you are going to generate an LF sweep with the fol-
lowing settings:
• frequency range from 10 to 60 kHz, at a level of 500 mV
• six sweep points and 50 ms dwell time at each point
Refer to the following procedures to configure and generate
an LF sweep:
1
Press
hardkey
2
Press
hardkey
3
Press
Step Sweep
softkey
4
Press
LF Start > 10 > kHz
5
Press
LF Stop > 60 > kHz
6
Press
# Points > 6 > Enter
7
Press
Step Dwell > 50 > ms
8
Press
Return > Sweep State > LF
This initiates the LF sweep output immediately via the
LF
OUT
connector. A
SWEEP
annunciator displays for the
duration of the sweep.
For more information about the sweep settings, please refer
to
NOTE
Pressing
sets the sweep/point trigger to Immediate and the LF
output amplitude to 500 mV. The following procedures leave out the steps
that set up the sweep/point trigger and the LF output amplitude.
Preset
Preset
Sweep