Vertical Controls
3
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
69
For certain active probes, such as InfiniiMax probes, the oscilloscope can
accurately calibrate its analog channels for the probe. When you connect a
probe that can be calibrated, the
Calibrate Probe
softkey in the Channel
Probe Menu becomes active.
To calibrate one of these probes:
1
First, plug your probe into one of the oscilloscope channels.
This could be, for example, an InfiniiMax probe amplifier/probe head
with attenuators attached.
2
Connect the probe to the left side, Demo 2, Probe Comp terminal, and
the probe ground to the ground terminal.
3
Press the Channel on/off key to turn the channel on (if the channel is
off).
4
In the Channel Menu, press the
Probe
softkey.
5
In the Channel Probe Menu, the second softkey from the left is for
specifying your probe head (and attenuation). Repeatedly press this
softkey until the probe head selection matches the attenuator you are
using.
The choices are:
•
10:1 single- ended browser (no attenuator).
•
10:1 differential browser (no attenuator).
•
10:1 (+6 dB Atten) single- ended browser.
•
10:1 (+6 dB Atten) differential browser.
•
10:1 (+12 dB Atten) single- ended browser.
•
10:1 (+12 dB Atten) differential browser.
•
10:1 (+20 dB Atten) single- ended browser.
•
10:1 (+20 dB Atten) differential browser.
6
Press the
Calibrate Probe
softkey and follow the instructions on the
display.
N O T E
When calibrating a differential probe, connect the positive lead to the Probe Comp terminal
and the negative lead to the ground terminal. You may need to connect an alligator clip to
the ground lug to allow a differential probe to span between the Probe Comp test point and
ground. A good ground connection ensures the most accurate probe calibration.
Содержание InfiniiVision 3000 3024A MSO-X
Страница 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Страница 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Страница 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Страница 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Страница 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Страница 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Страница 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Страница 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Страница 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Страница 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Страница 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Страница 278: ...278 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 17 Save Recall Setups Screens Data ...
Страница 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...
Страница 360: ...360 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 23 FlexRay Triggering and Serial Decode ...
Страница 380: ...380 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 24 I2C SPI Triggering and Serial Decode ...