24
I2C/SPI Triggering and Serial Decode
4
Press the
Mode
softkey; then, select
I2C
trigger type.
5
Press the
Signals
softkey to open the I
2
C Signals Menu.
6
For both the SCL (serial clock) and SDA (serial data) signals:
a
Connect an oscilloscope channel to the signal in the device under
test.
b
Press the
SCL
or
SDA
softkey; then, turn the Entry knob to select the
channel for the signal.
c
Press the corresponding
Threshold
softkey; then, turn the Entry knob
to select the signal threshold voltage level.
The threshold voltage level is used in decoding, and it will become
the trigger level when the trigger type is set to the selected serial
decode slot.
Data must be stable during the whole high clock cycle or it will be
interpreted as a start or stop condition (data transitioning while the
clock is high).
The SCL and SDA labels for the source channels are automatically set.
I2C Triggering
To set up the oscilloscope to capture I2C signals, see
"Setup for I2C
Signals"
on page 361.
After the oscilloscope has been set up to capture I2C signals, you can
trigger on a stop/start condition, a restart, a missing acknowledge, an
EEPROM data read, or on a read/write frame with a specific device
address and data value.
1
Press
[Trigger]
; then, select the
I2C
trigger type.
2
Press
[Trigger]
.
Содержание InfiniiVision 3000 3024A MSO-X
Страница 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Страница 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Страница 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Страница 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Страница 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Страница 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Страница 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Страница 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Страница 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Страница 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Страница 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Страница 278: ...278 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 17 Save Recall Setups Screens Data ...
Страница 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...
Страница 360: ...360 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 23 FlexRay Triggering and Serial Decode ...
Страница 380: ...380 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 24 I2C SPI Triggering and Serial Decode ...