I2S Triggering and Serial Decode
25
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
385
3
Press the
Trigger Setup
softkey to open the I
2
S Trigger Setup Menu.
4
Press the
Audio
softkey; then, turn the Entry knob to choose to trigger
on
Left
channel events,
Right
channel events, or events that occur on
Either
channel.
5
Press the
Trigger
softkey and choose a qualifier:
•
Equal
— triggers on the specified audio channel's data word when it
equals the specified word.
•
Not equal
— triggers on any word other than the specified word.
•
Less than
— triggers when the channel's data word is less than the
specified value.
•
Greater than
— triggers when the channel's data word is greater than
the specified value.
•
In Range
— enter upper and lower values to specify the range in
which to trigger.
•
Out of Range
— enter upper and lower values to specify range in which
trigger will not occur.
•
Increasing value
— triggers when the data value is increasing over time
and the specified value is met or exceeded. Set
Trigger >=
to the data
value that must be reached. Set
Armed <=
to the value to which the
data must fall before the trigger circuit is re- armed (ready to trigger
again). These settings are made in the current menu when
Base
is
Decimal
or in the Bits submenu when the
Base
is
Binary
. The Armed
control reduces triggers due to noise.
This trigger condition is best understood when the digital data
transferred over the I2S bus is considered in terms of representing
an analog waveform. The figure below shows a plot of sample data
transmitted over an I2S bus for one channel. In this example, the
oscilloscope will trigger at the 2 points shown since there are two
instances in which the data increases from a value below (or at) the
"Armed" value to a value greater than (or equal to) the specified
"Trigger" value.
Содержание InfiniiVision 3000 3024A MSO-X
Страница 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Страница 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Страница 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Страница 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Страница 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Страница 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Страница 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Страница 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Страница 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Страница 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Страница 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Страница 278: ...278 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 17 Save Recall Setups Screens Data ...
Страница 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...
Страница 360: ...360 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 23 FlexRay Triggering and Serial Decode ...
Страница 380: ...380 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 24 I2C SPI Triggering and Serial Decode ...