Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
24
I2C/SPI Triggering and Serial Decode
9
Press the
CS
softkey to open the SPI Chip Select Menu.
In the SPI Chip Select Menu:
a
Press the
Frame by
softkey to select a framing signal that the
oscilloscope will use for determining which clock edge is the first
clock edge in the serial stream.
You can set the oscilloscope to trigger during a high chip select (
CS
),
a low chip select (
~CS
), or after a
Timeout
period during which the
clock signal has been idle.
•
If the framing signal is set to
CS
(or
~CS
), the first clock edge as
defined, rising or falling, seen after the
CS
(or
~CS
) signal
transitions from low to high (or high to low) is the first clock in
the serial stream.
Chip Select
— Press the
CS
or
~CS
softkey; then, turn the Entry
knob to select the channel that is connected to the SPI frame line.
The label (~CS or CS) for the source channel is automatically set.
The data pattern and the clock transition must occur during the
time when the framing signal is valid. The framing signal must be
valid for the entire data pattern.
•
If the framing signal is set to
Timeout
, the oscilloscope generates
it's own internal framing signal after it sees inactivity on the serial
clock line.
Содержание InfiniiVision 3000 3024A MSO-X
Страница 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Страница 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Страница 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Страница 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Страница 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Страница 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Страница 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Страница 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Страница 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Страница 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Страница 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Страница 278: ...278 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 17 Save Recall Setups Screens Data ...
Страница 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...
Страница 360: ...360 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 23 FlexRay Triggering and Serial Decode ...
Страница 380: ...380 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 24 I2C SPI Triggering and Serial Decode ...