Mask Testing
15
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
245
To manually modify a mask file
You can manually modify a mask file that you created using the Automask
function.
1
Follow the steps 1- 7 in
"To create a mask from a "golden" waveform
(Automask)"
on page 239. Do not clear the mask after creating it.
2
Attach a USB mass storage device to the oscilloscope.
3
Press the
[Save/Recall]
key.
4
Press the
Save
softkey.
5
Press the
Format
softkey and select
Mask
.
6
Press the second softkey and select a destination folder on your USB
mass storage device.
7
Press the
Press to Save
softkey. This creates an ASCII text file that
describes the mask.
8
Remove the USB mass storage device and connect it to a PC.
9
Open the .msk file your created using a text editor (such as Wordpad).
10
Edit, save, and close the file.
The mask file contains the following sections:
•
Mask File Identifier.
•
Mask Title.
•
Mask Violation Regions.
Reset Statistics
Note that statistics are also reset when:
• Mask Test is switched on after being switched off.
• Clear Mask softkey is pressed.
• An Automask is created.
Additionally, the accumulated time counter is reset whenever the
oscilloscope is run after the acquisition was stopped.
Transparent
Enable Transparent mode to write measurement values and statistics on the
screen without a background. Disable the Transparent mode to show them
with a gray background. The Transparent setting affects mask test statistics,
measurement statistics, and reference waveform information display.
Clear Display
Clears acquisition data from the oscilloscope display.
Содержание InfiniiVision 3000 3024A MSO-X
Страница 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Страница 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Страница 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Страница 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Страница 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Страница 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Страница 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Страница 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Страница 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Страница 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Страница 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Страница 278: ...278 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 17 Save Recall Setups Screens Data ...
Страница 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...
Страница 360: ...360 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 23 FlexRay Triggering and Serial Decode ...
Страница 380: ...380 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 24 I2C SPI Triggering and Serial Decode ...