![Agilent Technologies InfiniiVision 3000 3024A MSO-X Скачать руководство пользователя страница 180](http://html1.mh-extra.com/html/agilent-technologies/infiniivision-3000-3024a-mso-x/infiniivision-3000-3024a-mso-x_user-manual_2868679180.webp)
180
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
11
Trigger Mode/Coupling
•
"To position the time reference (left, center, right)"
on page 56
To select the trigger coupling
1
Press the
[Mode/Coupling]
key.
2
In the Trigger Mode and Coupling Menu, press the
Coupling
softkey;
then, turn the Entry knob to select:
•
DC
coupling — allows DC and AC signals into the trigger path.
•
AC
coupling — places a 10 Hz high- pass filter in the trigger path
removing any DC offset voltage from the trigger waveform.
The high- pass filter in the External Trigger input path is 50 Hz for
all models.
Use AC coupling to get a stable edge trigger when your waveform has
a large DC offset.
•
LF
(low frequency)
Reject
coupling — adds a high- pass filter with the
3- dB point at 50 kHz in series with the trigger waveform.
Low frequency reject removes any unwanted low frequency
components from a trigger waveform, such as power line frequencies,
etc., that can interfere with proper triggering.
Use
LF Reject
coupling to get a stable edge trigger when your
waveform has low frequency noise.
•
TV
coupling — is normally grayed- out, but is automatically selected
when TV trigger is enabled in the Trigger Menu.
Pass
Band
50 kHz
0 dB
3 dB down point
DC
Содержание InfiniiVision 3000 3024A MSO-X
Страница 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Страница 22: ...22 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide ...
Страница 122: ...122 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Страница 128: ...128 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Страница 134: ...134 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Страница 140: ...140 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 9 Labels ...
Страница 176: ...176 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 10 Triggers ...
Страница 184: ...184 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Страница 202: ...202 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control ...
Страница 212: ...212 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors ...
Страница 262: ...262 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Waveform Generator ...
Страница 278: ...278 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 17 Save Recall Setups Screens Data ...
Страница 300: ...300 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Utility Settings ...
Страница 360: ...360 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 23 FlexRay Triggering and Serial Decode ...
Страница 380: ...380 Agilent InfiniiVision 3000 X Series Oscilloscopes User s Guide 24 I2C SPI Triggering and Serial Decode ...