Acquire Commands
:ACQuire:COMPlete
Command
:ACQuire:COMPlete <percent>
The :ACQuire:COMPlete command specifies how many of the data point
storage bins (time buckets) in the waveform record must contain a
waveform sample before a measurement will be made. For example, if the
command :ACQuire:COMPlete 60 has been sent, 60% of the storage bins in
the waveform record must contain a waveform data sample before a
measurement is made.
•
If :ACQuire:AVERage is set to OFF, the oscilloscope only needs one
value per time bucket for that time bucket to be considered full.
•
If :ACQuire:AVERage is set to ON, each time bucket must have
n
hits
for it to be considered full, where
n
is the value set by
:ACQuire:AVERage:COUNt.
Due to the nature of real time acquisition, 100% of the waveform record
bins are filled after each trigger event, and all of the previous data in the
record is replaced by new data when :ACQuire:AVERage is off. Hence, the
complete mode really has no effect, and the behavior of the oscilloscope is
the same as when the completion criteria is set to 100% (this is the same
as in PDETect mode). When :ACQuire:AVERage is on, all of the previous
data in the record is replaced by new data.
The range of the :ACQuire:COMPlete command is 0 to 100 and indicates
the percentage of time buckets that must be full before the acquisition is
considered complete. If the complete value is set to 100%, all time buckets
must contain data for the acquisition to be considered complete. If the
complete value is set to 0, then one acquisition cycle will take place.
Completion is set by default setup or *RST to 90%. Autoscale changes it to
100%.
<percent>
An integer, 0 to 100, representing the percentage of storage bins (time
buckets) that must be full before an acquisition is considered complete.
Example
This example sets the completion criteria for the next acquisition to 90%.
myScope.WriteString ":ACQUIRE:COMPLETE 90"
Query
:ACQuire:COMPlete?
The :ACQuire:COMPlete? query returns the completion criteria.
Returned Format
[:ACQuire:COMPlete] <percent><NL>
<percent>
An integer, 0 to 100, representing the percentage of time buckets that
must be full before an acquisition is considered complete.
Содержание Infiniium 90000 Q-Series
Страница 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Страница 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Страница 86: ...86 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 3 Introduction to Programming...
Страница 146: ...146 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 8 Programming Conventions...
Страница 190: ...190 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 11 Calibration Commands...
Страница 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Страница 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Страница 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Страница 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Страница 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Страница 404: ...404 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 20 InfiniiScan ISCan Commands...
Страница 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Страница 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Страница 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Страница 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Страница 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Страница 804: ...804 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 30 Serial Data Equalization Commands...
Страница 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Страница 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Страница 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Страница 1000: ...1000 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 35 Waveform Memory Commands...
Страница 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Страница 1109: ...Sample Programs 38 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 1109 print End of program...
Страница 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Страница 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...