Status Reporting
6
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
109
The next program prints "0x84" and clears bit 6 (RQS) of the Status Byte
Register. The difference in the decimal value between this example and the
previous one is the value of bit 6 (weight = 64). Bit 6 is set when the first
enabled summary bit is set, and is cleared when the Status Byte Register
is read by the serial poll command.
Example 2
The following example uses the resource session object's ReadSTB method
to read the contents of the oscilloscope's Status Byte Register.
varStbValue = myScope.IO.ReadSTB
Debug.Print "Status Byte Register, Serial Poll: 0x" + Hex(varStbValue)
N O T E
Use Serial Polling to Read the Status Byte Register
Serial polling is the preferred method to read the contents of the Status Byte Register
because it resets bit 6 and allows the next enabled event that occurs to generate a new
SRQ interrupt.
Содержание Infiniium 90000 Q-Series
Страница 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Страница 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Страница 86: ...86 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 3 Introduction to Programming...
Страница 146: ...146 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 8 Programming Conventions...
Страница 190: ...190 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 11 Calibration Commands...
Страница 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Страница 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Страница 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Страница 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Страница 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Страница 404: ...404 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 20 InfiniiScan ISCan Commands...
Страница 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Страница 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Страница 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Страница 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Страница 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Страница 804: ...804 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 30 Serial Data Equalization Commands...
Страница 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Страница 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Страница 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Страница 1000: ...1000 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 35 Waveform Memory Commands...
Страница 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Страница 1109: ...Sample Programs 38 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 1109 print End of program...
Страница 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Страница 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...