Mask Test Commands
:MTESt:IMPedance
Command
:MTESt:IMPedance {NONE | IMP75 | IMP100 | IMP110 | IMP120}
The :MTESt:IMPedance command sets the desired probe impedance of the
channel being used for mask testing. This impedance value is used when
starting a mask test to determine whether or not the correct Infiniium
probe is connected and in the case of the E2621A if the switch is set to
the correct impedance value.
Infiniium has an AutoProbe interface that detects probes that have Probe
ID resistors. If one of these probes is connected to the channel being mask
tested and is not the correct probe for the selected impedance, a warning
dialog box appears when the mask test is started from the human
interface.
This command is meant to be used in the setup section of a mask file.
NONE
Disables the probe impedance check.
IMP75
Enables the probe impedance check for the E2622A probe.
IMP100
Enables the probe impedance check for the E2621A probe with the switch
set to the 100 ohm position.
IMP110
Enables the probe impedance check for the E2621A probe with the switch
set to the 110 ohm position.
IMP120
Enables the probe impedance check for the E2621A probe with the switch
set to the 120 ohm position.
Example
This example sets the probe impedance of the channel being used for mask
testing to 100 ohms.
myScope.WriteString ":MTEST:IMPEDANCE IMP100"
Query
:MTESt:IMPedance?
The :MTESt:IMPedance? query returns the current value of the mask test
impedance.
Returned Format
[:MTESt:IMPedance] {NONE | IMP75 | IMP100 | IMP110 | IMP120}<NL>
Example
This example returns the current value of the mask test impedance and
prints the result to the computer screen.
myScope.WriteString ":SYSTEM:HEADER OFF"
' Response headers off.
myScope.WriteString ":MTEST:IMPEDANCE?"
strImpedance = myScope.ReadString
Debug.Print strImpedance
Содержание Infiniium 90000 Q-Series
Страница 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Страница 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Страница 86: ...86 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 3 Introduction to Programming...
Страница 146: ...146 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 8 Programming Conventions...
Страница 190: ...190 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 11 Calibration Commands...
Страница 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Страница 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Страница 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Страница 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Страница 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Страница 404: ...404 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 20 InfiniiScan ISCan Commands...
Страница 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Страница 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Страница 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Страница 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Страница 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Страница 804: ...804 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 30 Serial Data Equalization Commands...
Страница 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Страница 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Страница 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Страница 1000: ...1000 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 35 Waveform Memory Commands...
Страница 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Страница 1109: ...Sample Programs 38 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 1109 print End of program...
Страница 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Страница 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...