480
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
25
Measure Commands
•
For a rise time measurement, the leading (positive- going) edge of the
waveform must be displayed.
•
For a fall time measurement, the trailing (negative- going) edge of the
waveform must be displayed.
In jitter mode with jitter statistics enabled, measurements are made on all
data regardless of what is on screen.
User-Defined
Thresholds
If you choose to set user- defined thresholds, they must be set before
actually sending the measurement command or query.
Measurement
Error
If a measurement cannot be made because of a lack of data, because the
source waveform is not displayed, the requested measurement is not
possible (for example, a period measurement on an FFT waveform), or for
some other reason, the following results are returned:
•
9.37 is returned as the measurement result.
•
If SENDvalid is ON, the error code is also returned as well as the
questionable value.
Making
Measurements
If more than one period, edge, or pulse is displayed, time measurements
are made on the first, left- most portion of the displayed waveform.
When any of the defined measurements are requested, the oscilloscope
first determines the top (100%) and base (0%) voltages of the waveform.
From this information, the oscilloscope determines the other important
voltage values (10%, 90%, and 50% voltage values) for making
measurements.
The 10% and 90% voltage values are used in the rise time and fall time
measurements when standard thresholds are selected. The 50% voltage
value is used for measuring frequency, period, pulse width, and duty cycle
with standard thresholds selected.
You can also make measurements using user- defined thresholds instead of
the standard thresholds.
When the command form of a measurement is used, the oscilloscope is
placed in the continuous measurement mode. The measurement result will
be displayed on the front panel. There may be a maximum of 5
measurements running continuously. Use the SCRatch command to turn
off the measurements.
When the query form of the measurement is used, the measurement is
made one time, and the measurement result is returned.
•
If the current acquisition is complete, the current acquisition is
measured and the result is returned.
•
If the current acquisition is incomplete and the oscilloscope is running,
acquisitions will continue to occur until the acquisition is complete. The
acquisition will then be measured and the result returned.
Содержание Infiniium 90000 Q-Series
Страница 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Страница 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Страница 86: ...86 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 3 Introduction to Programming...
Страница 146: ...146 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 8 Programming Conventions...
Страница 190: ...190 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 11 Calibration Commands...
Страница 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Страница 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Страница 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Страница 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Страница 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Страница 404: ...404 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 20 InfiniiScan ISCan Commands...
Страница 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Страница 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Страница 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Страница 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Страница 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Страница 804: ...804 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 30 Serial Data Equalization Commands...
Страница 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Страница 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Страница 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Страница 1000: ...1000 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 35 Waveform Memory Commands...
Страница 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Страница 1109: ...Sample Programs 38 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 1109 print End of program...
Страница 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Страница 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...