Channel Commands
:CHANnel<N>:PROBe:PRECprobe:ZSRC
Command
:CHANnel<N>:PROBe:PRECprobe:ZSRC {VIN | {VSRC, <impedance>}
| {VSRC, <file_string>}}
The :CHANnel<N>:PROBe:PRECprobe:ZSRC command specifies how
PrecisionProbe characterizes the time domain and frequency domain
response.
<N>
An integer, 1- 4.
VIN
Selects the VOut/Vin probe transfer function (which characterizes the
output of the probe as a function of the input at the probe tips).
Defining the response this way lets you evaluate the probe's accuracy in
reproducing the actual signal present in your system with the probe
attached. This correction is what you would see with a real band limited
probe that has finite input impedance. PrecisionProbe corrects the
"VOut/Vin" response to be flat with frequency and phase to your defined
bandwidth limit. It does not de- embed the loading effects of the probe.
(Agilent's probe corrections are typically defined using Vout/Vin.)
VSRC,
<impedance>
Selects the VOut/VSrc estimate of probed system response (which corrects
the probe as "what would be there if the probe were not present"), and
specifies a constant (Zo/2) value (in ohms) as the system source
impedance.
One drawback of defining the probe's response in this manner is that if
the probe's loading causes your circuit to lose some timing or amplitude
margin, you probably want to know that when you make a measurement.
VOut/VSource compensation will hide these effects from you. However, this
method can be effective if probing at the transmitter.
VSRC,
<file_string>
Selects the VOut/VSrc estimate of probed system response (which corrects
the probe as "what would be there if the probe were not present"), and
names an S- parameter file whose S11 is used to specify the system source
impedance.
Example
This example, for channel 1, tells PrecisionProbe to use the VOut/VSrc
characterization and to get the system source impedance from S11 in the
"foo.s2p" S- parameter file.
myScope.WriteString ":CHANNEL1:PROBE:PRECprobe:ZSRC VSRC, "foo.s2p""
Query
:CHANnel<N>:PROBe:PRECprobe:ZSRC?
The :CHANnel<N>:PROBe:PRECprobe:ZSRC? query returns the current
settings for PrecisionProbe time domain and frequency domain response
characterization.
Returned Format
[:CHANnel<N>:PROBe:PRECprobe:ZSRC] {VIN | {VSRC, <impedance>}
| {VSRC, <file_string>}}<NL>
Содержание Infiniium 90000 Q-Series
Страница 1: ...s1 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference...
Страница 50: ...50 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 2 Setting Up...
Страница 86: ...86 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 3 Introduction to Programming...
Страница 146: ...146 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 8 Programming Conventions...
Страница 190: ...190 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 11 Calibration Commands...
Страница 274: ...274 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 14 Digital Commands...
Страница 292: ...292 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 15 Disk Commands...
Страница 318: ...318 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 16 Display Commands...
Страница 364: ...364 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 17 Function Commands...
Страница 382: ...382 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 19 Histogram Commands...
Страница 404: ...404 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 20 InfiniiScan ISCan Commands...
Страница 412: ...412 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 21 Limit Test Commands...
Страница 416: ...416 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 22 Lister Commands...
Страница 696: ...696 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 25 Measure Commands...
Страница 702: ...702 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 26 Pod Commands...
Страница 760: ...760 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 29 Self Test Commands...
Страница 804: ...804 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 30 Serial Data Equalization Commands...
Страница 818: ...818 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 31 System Commands...
Страница 830: ...830 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 32 Time Base Commands...
Страница 930: ...930 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 33 Trigger Commands...
Страница 1000: ...1000 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 35 Waveform Memory Commands...
Страница 1018: ...1018 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 37 Error Messages...
Страница 1109: ...Sample Programs 38 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 1109 print End of program...
Страница 1150: ...1150 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference 38 Sample Programs...
Страница 1186: ...1186 Agilent Infiniium 90000 Series Oscilloscopes Programmer s Reference Index...