Vacuum Products Division
Manual 699910005, Hot Ion Source Replacement
Page 17
3.
Using an extended length M5 Allen wrench, remove the four screws holding the rear plastic cover
(Figure 3-1).
4.
Detach the rear plastic cover from the unit.
Figure 3-2. Fan assembly
5.
Disconnect the fan cable and remove the fan assembly by unfastening two wing nuts and one M4
socket head cap screw (Figure 3-2).
Figure 3-3. Cable locations
CAUTION
Static sensitive device, ensure that personnel are properly grounded
before proceeding.
6.
Disconnect the spectrometer ground cable, the preamp signal cable and the ion source cable from
the spectrometer (Figure 3-3).
NOTE
Cables not shown for clarity.