Vacuum Products Division
Manual 699910005, Hot Ion Source Replacement
Page 23
19.
Connect the collimator wire to the collimator plate using an M 1.5 screw. See Figure 3-8.
20.
Connect the 4 coupler screws to their respective filaments.
Figure 3-8. Collimator wire connection
21.
Insert the high voltage ion source wire (brownish-orange wire in rendering) into the ion chamber
and tighten the M1.6 socket head screw.
NOTE
Pull lightly with small pliers on the high voltage ion source wire
(brownish-orange) to ensure it is connected securely and will not pull out.
Figure 3-9. Ion chamber and extractor plate
Collimator
screw
Ion
chamber
Extractor
plate