Vacuum Products Division
Manual 699910005, Hot Ion Source Replacement
Page 20
CAUTION
Wear non-powdered, ESD-safe Nitride or equivalent gloves (not included
in kit) to prevent skin oils from getting on vacuum surfaces.
Refer to Figure 2-3 for disassembly of the spectrometer header.
8.
Loosen screws (10X) and remove spectrometer header from spectrometer body. The vacuum
system vents to atmosphere as the screws are loosened. Retain the socket head cap screws and
Belleville washers. Maintain correct orientation of Belleville (conical) washers when removing.
9.
Remove the spectrometer header. Rest upside-down on the four posts sticking up from the printed
circuit board.
10.
Remove metal gasket and discard. Do not scratch the mating surface when removing the gasket.
To prevent scratching of any sealing surface, do not use metal tools for this task. Do not attempt to
reuse the gasket.
11.
Examine the spectrometer cavity for discolored areas. If the inside of the spectrometer body
appears dirty, clean it with Vac-u-solv spectrometer cleaning kit (Agilent part number 670029096).