Repair and Replacement Procedures
Removing and Replacing the Measurement Group Assemblies
Replacing the A5 System Board
1.
Separate the front and rear cases. See
“Separating the Front and Rear Cases” on page 7-10
.
2. Remove all measurement group assemblies as one unit. Refer to
Group Assemblies as One Unit” on page 7-27
3. Remove the I/O side panel. Refer to
“Replacing the I/O Side Panel, Doors, and Gasket” on page
4. Disconnect cable W3 and W4 from the A5 System board.
5. Refer to
. Remove the RTC board. Refer to
“Replacing the Real Time Clock (RTC)
Battery Located on the A5 System Board Shield” on page 7-31
6. Remove the System board shield by removing the nineteen screws (item 2). Discard the screws, new
ones are provided in the
7. Refer to
. Remove the four screws (item 2) that secure the A6 SOM board (item 1) to the
A5 System board. Discard these screws; new ones are provided in the
8. Unplug the A6 SOM board from the A5 System board.
9. Refer to
. Remove the two screws (item 1) from the A5 System board, lift it up from the
RF bottom shield, and unplug it from the A4 RF board connector. Discard the screws, new ones are
provided in the
.
10. Refer to
. Plug in the new A5 System board to the A4 RF board, over the new gasket, and
aligned with the three dowel pins in the RF bottom shield. Make sure the bullets are straight up when
connecting the System and RF boards.
. Replace the SOM board onto the System board. Install four new screws (item 2)
from the
and tighten them in the torque sequence indicated by the circled
numbers (1 - 2 - 3 - 4) in the illustration.
12. Refer to
. Place the System shield onto the System board by aligning the dowel pins.
Replace the nineteen shield screws. Torque these screws (plus the two System board screws, #13 and
#21) in the sequence shown by the circled numbers (1 - 21). Repeat this torque sequence one
additional time. Torque to 12 in-lbs.
13. Place the RTC board onto the System board shield and secure with the two screws. Torque to 9 in-lbs.
14. Remove the System board to FPIB cable (W14) from the old A5 System board and install it on the
new board.
15. Reinstall the I/O side panel. Refer to
“Replacing the I/O Side Panel, Doors, and Gasket” on page
16. Perform the steps under
Содержание FieldFox MW N9913A
Страница 8: ...Contents 4 Service Guide N9927 90003 Contents ...
Страница 9: ...N9927 90003 1 1 1 Safety and Regulatory Information ...
Страница 16: ...1 8 N9927 90003 Safety and Regulatory Information FieldFox Analyzer Regulatory Information ...
Страница 17: ...N9927 90003 2 1 2 General Product Information ...
Страница 26: ...2 10 N9927 90003 General Product Information FieldFox Analyzer Agilent Support Services and Assistance ...
Страница 27: ...N9927 90003 3 1 3 Tests and Adjustments ...
Страница 37: ...N9927 90003 4 1 4 Troubleshooting ...
Страница 50: ...4 14 N9927 90003 Troubleshooting FieldFox Analyzer Measurement Group Troubleshooting ...
Страница 51: ...N9927 90003 5 1 5 Theory of Operation ...
Страница 58: ...5 8 N9927 90003 Theory of Operation FieldFox Analyzer Measurement Group Figure 5 2 Simplified RF Block Diagram ...
Страница 81: ...N9927 90003 6 1 6 Replaceable Parts ...
Страница 95: ...N9927 90003 7 1 7 Repair and Replacement Procedures ...
Страница 140: ...7 46 N9927 90003 Repair and Replacement Procedures FieldFox Analyzer Post Repair Procedures ...
Страница 141: ...N9927 90003 8 1 8 General Purpose Maintenance and Service Procedures ...
Страница 149: ...N9927 90003 Appendix 1 Appendix A ...
Страница 156: ...Index 4 N9927 90003 Index ...