N9927-90003
5-21
FieldFox Analyzer
Theory of Operation
Grounding and Power
Battery Charging Management
The battery pack determines when the battery needs charging and requests a charge over the Smart Batt
Bus.
The Battery Charger and Manager circuit on the A5 System board responds by charging the battery if
there is sufficient voltage on the Ext DC line. A minimum of approximately 15 volts is required to charge
the battery.
The System Monitor on the A5 System board monitors the battery’s temperature and, if too high or too
low, disables charging of the battery.
How to tell if the battery is charging:
•
If the analyzer is On, the lightening bolt is displayed in the battery icon.
•
If the analyzer is Off, the LED below the On/Off Button slowly flashes orange with increasing
intensity.
•
If the analyzer is in Standby, turn the analyzer on and observe the battery icon.
The battery charges much faster up to 70% of its full charge than it does above 70% of full charge.
Battery Life
When a battery is new, an instrument should operate for about four hours on a fully charged battery. The
actual time will depend on operating mode and temperature.
As a battery goes through charge/discharge cycles, its capacity will decrease. A typical battery will have
more than 70% of its initial capacity after 200 full discharge cycles when operating at 20 to 30
°
C
ambient temperature.
Содержание FieldFox MW N9913A
Страница 8: ...Contents 4 Service Guide N9927 90003 Contents ...
Страница 9: ...N9927 90003 1 1 1 Safety and Regulatory Information ...
Страница 16: ...1 8 N9927 90003 Safety and Regulatory Information FieldFox Analyzer Regulatory Information ...
Страница 17: ...N9927 90003 2 1 2 General Product Information ...
Страница 26: ...2 10 N9927 90003 General Product Information FieldFox Analyzer Agilent Support Services and Assistance ...
Страница 27: ...N9927 90003 3 1 3 Tests and Adjustments ...
Страница 37: ...N9927 90003 4 1 4 Troubleshooting ...
Страница 50: ...4 14 N9927 90003 Troubleshooting FieldFox Analyzer Measurement Group Troubleshooting ...
Страница 51: ...N9927 90003 5 1 5 Theory of Operation ...
Страница 58: ...5 8 N9927 90003 Theory of Operation FieldFox Analyzer Measurement Group Figure 5 2 Simplified RF Block Diagram ...
Страница 81: ...N9927 90003 6 1 6 Replaceable Parts ...
Страница 95: ...N9927 90003 7 1 7 Repair and Replacement Procedures ...
Страница 140: ...7 46 N9927 90003 Repair and Replacement Procedures FieldFox Analyzer Post Repair Procedures ...
Страница 141: ...N9927 90003 8 1 8 General Purpose Maintenance and Service Procedures ...
Страница 149: ...N9927 90003 Appendix 1 Appendix A ...
Страница 156: ...Index 4 N9927 90003 Index ...