Tests and Adjustments
FieldFox Analyzer
Before You Begin
Review the Principles of Connector Care
Proper connector care and connection techniques are critical for accurate and repeatable measurements.
Refer to Table 3-1
for tips on connector care.
Prior to making connections to your analyzer, carefully review the information about inspecting,
cleaning, and gaging connectors. Refer to the calibration kit documentation for detailed connector care
information.
For course numbers about additional connector care instruction, contact Agilent Technologies. Refer to
“Contacting Agilent” on page 2-9
.
Table 3-1Connector Care Quick Reference Guide
Handling and Storage
Do
• Keep connectors clean
Do Not
• Touch mating-plane surfaces
• Extend sleeve or connector nut
• Set connectors contact-end down
• Use plastic end-caps during storage
• Store connectors or adapters loose
Visual Inspection
Do
• Inspect all connectors carefully
Do Not
• Use a damaged connector - ever
• Look for metal particles, scratches, and
dents
Connector Cleaning
Do
• Try compressed air first
Do Not
• Use any abrasives
• Use isopropyl alcohol
1
1. Cleaning connectors with alcohol shall only be done with the instrument’s power cord removed, and in
a well-ventilated area. Allow all residual alcohol moisture to evaporate, and the fumes to dissipate prior
to energizing the instrument.
• Get liquid into plastic support beads
• Clean connector threads
Gaging Connectors
Do
• Clean and zero the gage before use
Do Not
• Use an out-of-specification connector
• Use the correct gage type
• Use correct end of calibration block
• Gage all connectors before first use
Making Connections
Do
• Align connectors carefully
Do Not
• Apply bending force to connection
• Make preliminary connection contact
lightly
• Over tighten preliminary connection
• Turn only the connector nut
• Twist or screw any connection
• Use a torque wrench for final connection
• Tighten past torque wrench “break”
point
Содержание FieldFox MW N9913A
Страница 8: ...Contents 4 Service Guide N9927 90003 Contents ...
Страница 9: ...N9927 90003 1 1 1 Safety and Regulatory Information ...
Страница 16: ...1 8 N9927 90003 Safety and Regulatory Information FieldFox Analyzer Regulatory Information ...
Страница 17: ...N9927 90003 2 1 2 General Product Information ...
Страница 26: ...2 10 N9927 90003 General Product Information FieldFox Analyzer Agilent Support Services and Assistance ...
Страница 27: ...N9927 90003 3 1 3 Tests and Adjustments ...
Страница 37: ...N9927 90003 4 1 4 Troubleshooting ...
Страница 50: ...4 14 N9927 90003 Troubleshooting FieldFox Analyzer Measurement Group Troubleshooting ...
Страница 51: ...N9927 90003 5 1 5 Theory of Operation ...
Страница 58: ...5 8 N9927 90003 Theory of Operation FieldFox Analyzer Measurement Group Figure 5 2 Simplified RF Block Diagram ...
Страница 81: ...N9927 90003 6 1 6 Replaceable Parts ...
Страница 95: ...N9927 90003 7 1 7 Repair and Replacement Procedures ...
Страница 140: ...7 46 N9927 90003 Repair and Replacement Procedures FieldFox Analyzer Post Repair Procedures ...
Страница 141: ...N9927 90003 8 1 8 General Purpose Maintenance and Service Procedures ...
Страница 149: ...N9927 90003 Appendix 1 Appendix A ...
Страница 156: ...Index 4 N9927 90003 Index ...