N9927-90003
5-11
FieldFox Analyzer
Theory of Operation
Measurement Group
SA RF Section
An external RF input signal can be analyzed when connected to the Port 2 SA RF Input connector.
Before the input signal is mixed, the signal will go through the input attenuator and optional internal
preamplifier.
The input attenuator default setting from the factory is 10 dB. The input attenuator is used to control the
amount of power at the input of the 1st converter. Changing the input attenuator value can improve the
dynamic range of the instrument and reduce internally generated distortion products.
When the optional internal preamplifier is turned on, the displayed average noise level of the instrument
is improved. This occurs because the noise figure of the overall system is improved. Improving the
displayed average noise level of the system enables the user to analyze low level input signals.
The 1st, 2nd and 3rd converters are used to downconvert the input signal to a final I.F. of 33.75 MHz.
That I.F. is fed to the ADC on the A5, System board. This final I.F. on the output of the ADC can be
measured at the IF Out SMB (m) connector on the side of the instrument. Refer to
The frequency bands for the SA circuitry on the RF board are listed below:
Band
Instrument
Start
Stop
1
0.030
15.000
2
15.000
500.000
3
500.000
1366.250
4
1366.250
2600.000
5
2600.000
3900.000
6
3900.000
7500.000
7
7500.000
9500.000
8
9500.000
12300.000
9
12300.000
15700.000
10
15700.000
16300.000
11
16300.000
19500.000
12
19500.000
23000.000
13
23000.000
23734.000
14
23734.000
25500.000
15
25500.000
26500.000
Содержание FieldFox MW N9913A
Страница 8: ...Contents 4 Service Guide N9927 90003 Contents ...
Страница 9: ...N9927 90003 1 1 1 Safety and Regulatory Information ...
Страница 16: ...1 8 N9927 90003 Safety and Regulatory Information FieldFox Analyzer Regulatory Information ...
Страница 17: ...N9927 90003 2 1 2 General Product Information ...
Страница 26: ...2 10 N9927 90003 General Product Information FieldFox Analyzer Agilent Support Services and Assistance ...
Страница 27: ...N9927 90003 3 1 3 Tests and Adjustments ...
Страница 37: ...N9927 90003 4 1 4 Troubleshooting ...
Страница 50: ...4 14 N9927 90003 Troubleshooting FieldFox Analyzer Measurement Group Troubleshooting ...
Страница 51: ...N9927 90003 5 1 5 Theory of Operation ...
Страница 58: ...5 8 N9927 90003 Theory of Operation FieldFox Analyzer Measurement Group Figure 5 2 Simplified RF Block Diagram ...
Страница 81: ...N9927 90003 6 1 6 Replaceable Parts ...
Страница 95: ...N9927 90003 7 1 7 Repair and Replacement Procedures ...
Страница 140: ...7 46 N9927 90003 Repair and Replacement Procedures FieldFox Analyzer Post Repair Procedures ...
Страница 141: ...N9927 90003 8 1 8 General Purpose Maintenance and Service Procedures ...
Страница 149: ...N9927 90003 Appendix 1 Appendix A ...
Страница 156: ...Index 4 N9927 90003 Index ...