7-44
N9927-90003
Repair and Replacement Procedures
FieldFox Analyzer
Post-Repair Procedures
Post-replacement Procedure
1. Reinstall the measurement group assemblies into the front case. Refer to
Measurement Group Assemblies as One Unit” on page 7-42
2. Combine the front and rear cases. Refer to
“Combining the Front and Rear Cases” on page 7-18
Post-Repair Procedures
After the replacement of an assembly, you must perform the service procedures in the order listed in
Table 7-2.
Procedures referenced in this table are located in
Chapter 3, “Tests and Adjustments,”
unless
specified otherwise.
* Refer to the following manual checks and procedures for details.
Table 7-2 Related Service Procedures
Replaced Assembly
Adjustments, etc.
TME Tests or Other Procedures
Main battery
None needed.
None needed.
RPG
None needed.
Manual check of RPG operation.*
PORT 1 connector or
associated cable (W1)
replaced or disturbed.
Error Terms adjustment if applicable.
Error Terms test if applicable.
PORT 2 connector or
associated cable (W2)
replaced or disturbed.
Error Terms adjustment if applicable.
Error Terms test if applicable.
Absolute Amplitude Accuracy Test.
Total Abs Amp Accy LF Test.
Total Abs Amp Accy HF Test.
Front case (including
A1 keypad)
None needed.
Manual check of keyboard.*
A2 LCD
None needed.
Manual check of LCD.*
A3 front panel interface
board
None needed.
Manual check of keyboard.*
Check for even brightness in LCD
backlight.*
Check sound.*
A4 RF board
All TME based adjustments that apply to the
installed option configuration.
All TME based tests that apply to the
installed option configuration.
A5 System board
First, load the instrument serial number into
the A5 System board.*
All TME based adjustments that apply to the
installed option configuration.
All TME based tests that apply to the
installed option configuration.
A6 SOM board
Install firmware from SD card.*
Re-install options.*
All TME based adjustments that apply to the
installed option configuration.
Compare serial numbers.*
All tests that apply to the installed
option configuration.
A7 RTC Battery
None needed.
Check time and date.*
Содержание FieldFox MW N9913A
Страница 8: ...Contents 4 Service Guide N9927 90003 Contents ...
Страница 9: ...N9927 90003 1 1 1 Safety and Regulatory Information ...
Страница 16: ...1 8 N9927 90003 Safety and Regulatory Information FieldFox Analyzer Regulatory Information ...
Страница 17: ...N9927 90003 2 1 2 General Product Information ...
Страница 26: ...2 10 N9927 90003 General Product Information FieldFox Analyzer Agilent Support Services and Assistance ...
Страница 27: ...N9927 90003 3 1 3 Tests and Adjustments ...
Страница 37: ...N9927 90003 4 1 4 Troubleshooting ...
Страница 50: ...4 14 N9927 90003 Troubleshooting FieldFox Analyzer Measurement Group Troubleshooting ...
Страница 51: ...N9927 90003 5 1 5 Theory of Operation ...
Страница 58: ...5 8 N9927 90003 Theory of Operation FieldFox Analyzer Measurement Group Figure 5 2 Simplified RF Block Diagram ...
Страница 81: ...N9927 90003 6 1 6 Replaceable Parts ...
Страница 95: ...N9927 90003 7 1 7 Repair and Replacement Procedures ...
Страница 140: ...7 46 N9927 90003 Repair and Replacement Procedures FieldFox Analyzer Post Repair Procedures ...
Страница 141: ...N9927 90003 8 1 8 General Purpose Maintenance and Service Procedures ...
Страница 149: ...N9927 90003 Appendix 1 Appendix A ...
Страница 156: ...Index 4 N9927 90003 Index ...