8
Installation Note E8364-90022
Test Equipment Required for Post-Upgrade Adjustments
Step 1. Verify the Model, Serial, and Option Numbers of the Analyzer
Refer to the License Key Certificate supplied with the kit:
• Verify that the analyzer’s model and serial numbers match those on the certificate.
• Verify that the option number on the certificate matches the option described by this
installation note.
If the model, serial, or option number do not match the expected results, you will not be able
to install the option. If this is the case, contact Agilent for assistance. Refer to
.
Description
Agilent Part Number
Power meter
E4418B/E4419B
Power sensor, Type-N (100 kHz–4.2 GHz)
8482A
Power sensor, 2.4 mm (4.2 GHz–50 GHz)
8487A
Power sensor, V-Band (50 GHz–67 GHz)
a
a. This sensor and adapter must be calibrated together by the Agilent factory. The data supplied
are only valid as long as the sensor and adapter remain connected.
V8486A
Adapter, WR-15 to 1.85 mm (m)
V281B
Adapter, 1.85 mm (f) to 1.85 mm (f)
b
b. The 1.85 mm connector has the same ruggedness and is compatible with the 2.4 mm connec-
tor. 1.85 mm and 2.4 mm connectors and adapters may be used interchangeably.
85058-60114
Adapter, 3.5 mm (m) to 2.4 mm (f)
11901D
Adapter, 2.4 mm (f) to 2.4 mm (f)
11900B
Adapter, Type-N (f) to 2.4 mm (f)
11903B
Adapter, SMA (m) to SMA (f), right angle
1250-1249
Test cable, 2.4 mm (f) to 2.4 mm (f)
Test cable, 1.85 mm (f) to 1.85 mm (f)
85133C
N4697E
5/16-inch, open-end torque wrench (set to 10 in-lbs)
8710-1765
1/4-inch, open-end wrench, thin profile
8710-0510