Installation Note E8364-90022
17
Remove the A22 Switch/Splitter
1. Disconnect the stranded switch control cable (item
①
) from the A22 switch/splitter.
Remove this cable completely from the analyzer and discard it. It will be replaced later.
2. Using a 5/16-inch wrench, remove the five semirigid cables (item
②
). Discard these cables,
they will be replaced with new ones.
3. Using a T-10 TORX driver, remove the two mounting screws (item
③
). Retain the
mounting screws for reinstallation later.
CAUTION
Be careful not to damage the center pins of the semirigid cables. Some flexing of
the cables is necessary to remove the assembly. Do not over-bend them.
4. Remove the A22 switch/splitter from the analyzer. Observe the
CAUTION
above. Discard this
switch/splitter, it will be replaced with a new one.
Figure 9
A22 Switch/Splitter Removal