Installation Note E8364-90022
Install the New A18 MA 24
1. Position the new A18 MA 24 in the analyzer as shown. Observe the
CAUTION
below.
CAUTION
Be careful not to damage the center pins of the semirigid cables. Some flexing of
the cables is necessary when installing the assembly. Do not over-bend them.
2. Using a T-10 TORX driver, install the four mounting screws (item
③
), previously removed
from the old A18 MA 20, through the side of the analyzer.
3. Using a 5/16-inch torque wrench (set to 10 in-lbs), reconnect the two semirigid cables
(item
①
) to the A18 MA 24. Observe the
CAUTION
above.
4. Reconnect the ribbon cable (item
②
) to the A18 MA 24.
Figure 14
A18 MA 24 Installation