29
Install the New Cables
CAUTION
Be very careful with these cables with 1.85 mm connectors. They are extremely
delicate and can be easily damaged.
Using a 5/16-inch torque wrench set to 10 in-lbs, install the following new cables
in the
order listed
. These new parts are listed in
.
• W2
A21 SOMA 50 to A22 switch
• W3
A22 switch to A23 SOMA 70
• W4
A22 switch to A24 SOMA 70
• W8
A24 SOMA 70 to A29 channel R2 mixer
• W6
A24 SOMA 70 to A26 test port 2 coupler
• W10 A26 test port 2 coupler to A30 channel B mixer
• W7
A23 SOMA 70 to A28 channel R1 mixer
• W5
A23 SOMA 70 to A25 test port 1 coupler
• W9
A25 test port coupler 1 to A27 channel A mixer
• W38 A23 SOMA 70 to A16 test set motherboard J204 (R1)
• W39 A24 SOMA 70 to A16 test set motherboard J205 (R2)
•
①
Wrapped wire cable; A22 switch to A16 test set motherboard J500. Route this cable
under the cable clamp as shown.
•
②
Ribbon cable; A24 SOMA 70 to A16 test set motherboard J40 (
SOMA70_2
). Shape
and route this cable as shown.
•
③
Ribbon cable; A23 SOMA 70 to A16 test set motherboard J50 (
SOMA70_1
). Shape
and route this cable as shown.