70
Chapter 3
Optimizing Performance
Creating and Applying User Flatness Correction
Required Equipment
•
Agilent 83554A millimeter-wave source module
•
Agilent E4416A/17A/18B/19B power meter
•
Agilent R8486A power sensor
•
Agilent 8349B microwave amplifier (required for signal generators without Option 1EA)
•
GPIB interface cable
•
adapters and cables as required
NOTE
The equipment setups in
Figure 3-5
and
Figure 3-6
assume that the steps necessary to
correctly level the RF output have been followed. If you have questions about leveling with
a millimeter-wave source module, refer to
“To Level with a mm-Wave Source Module” on
page 63
.
Configure the Power Meter
1. Select SCPI as the remote language for the power meter.
2. Zero and calibrate the power sensor to the power meter.
3. Enter the appropriate power sensor calibration factors into the power meter as appropriate.
4. Enable the power meter’s cal factor array.
NOTE
For operating information on your particular power meter/sensor, refer to their operating
guides.
Содержание E8247C
Страница 10: ...Contents x ...
Страница 96: ...86 Chapter 4 Analog Modulation Configuring the LF Output ...
Страница 127: ...Chapter 5 117 Dual Arbitrary Waveform Generator Using Waveform Clipping Figure 5 16 Rectangular Clipping ...
Страница 128: ...118 Chapter 5 Dual Arbitrary Waveform Generator Using Waveform Clipping Figure 5 17 Reduction of Peak to Average Power ...
Страница 142: ...132 Chapter 6 Custom Arb Waveform Generator Working with Filters ...
Страница 178: ...168 Chapter 7 Custom Real Time I Q Baseband Working with Differential Data Encoding ...
Страница 198: ...188 Chapter 10 Troubleshooting ...
Страница 214: ...204 Chapter 10 Troubleshooting Returning a Signal Generator to Agilent Technologies ...