104
Chapter 5
Dual Arbitrary Waveform Generator
Using Waveform Markers
Using Waveform Markers
Waveform markers provide auxiliary output signals that are synchronized with a waveform segment. You
can place up to four markers on a waveform segment. However, only Marker 1 and Marker 2 can be placed
using the waveform player’s user interface (for more information, refer to
“Waveform Marker Concepts” on
page 108
).
Using markers, you can construct an output signal as a trigger to synchronize another instrument to a given
portion of a waveform. You can also place markers into a waveform sequence, either as the sequence is
being built or within an existing waveform sequence.
For instructions on verifying marker operation, see
“To Verify Marker Operation” on page 107
.
To Place a Marker at the First Point within a Waveform Segment
If you have not created a waveform segment, complete the steps in the previous sections,
“Generating the
First Waveform” on page 100
and
“Creating the First Waveform Segment” on page 100
.
1. Press
Mode
>
Dual ARB
>
Waveform Segments
.
2. Press
Load Store
.
3. Highlight a waveform segment (for example,
TTONE
).
4. Press
Waveform Utilities
>
Set Markers
>
Set Marker On First Point
.
This sets Marker 1 (selected by default) on the first point in the selected waveform segment.
To Place a Marker Across a Range of Points within a Waveform Segment
If you have not created a waveform segment, complete the steps in the previous sections,
“Generating the
First Waveform” on page 100
and
“Creating the First Waveform Segment” on page 100
.
1. Press
Mode
>
Dual ARB
>
Waveform Segments
.
2. Press
Load Store
.
3. Highlight a waveform segment (for example,
TTONE
).
4. Press
Waveform Utilities
>
Set Markers
>
Set Marker On Range Of Points
.
5. Press
First Mkr Point
>
10
>
Enter
.
6. Press
Last Mkr Point
>
163830
>
Enter
.
7. Press
Apply To Waveform
.
NOTE
The last marker point must be greater than or equal to the first marker point.
This activates Marker 1 (selected by default) from point 10 to point 163830 in the selected waveform
segment.
Содержание E8247C
Страница 10: ...Contents x ...
Страница 96: ...86 Chapter 4 Analog Modulation Configuring the LF Output ...
Страница 127: ...Chapter 5 117 Dual Arbitrary Waveform Generator Using Waveform Clipping Figure 5 16 Rectangular Clipping ...
Страница 128: ...118 Chapter 5 Dual Arbitrary Waveform Generator Using Waveform Clipping Figure 5 17 Reduction of Peak to Average Power ...
Страница 142: ...132 Chapter 6 Custom Arb Waveform Generator Working with Filters ...
Страница 178: ...168 Chapter 7 Custom Real Time I Q Baseband Working with Differential Data Encoding ...
Страница 198: ...188 Chapter 10 Troubleshooting ...
Страница 214: ...204 Chapter 10 Troubleshooting Returning a Signal Generator to Agilent Technologies ...