46
Chapter 2
Basic Operation
Configuring the RF Output
Figure 2-8
RS-232 Pin Configuration
2. Set up the slave PSG’s frequency and power settings.
By setting up the slave first, you avoid synchronization problems.
3. Set up the master PSG’s frequency, power, and sweep time settings.
The two PSGs can have different frequency and power settings for ramp sweep.
4. Set the slave PSG’s sweep time to match that of the master.
Sweep times must be the same for both PSGs.
5. Set the slave PSG to continuous triggering.
The slave must be set to continuous triggering, but the master can be set to any triggering mode.
6. On the slave PSG, press
Sweep/List
>
Sweep Type
>
Sweep Control
>
Slave
.
This sets the PSG to operate in slave mode.
7. On the master PSG, press
Sweep/List
>
Sweep Type
>
Sweep Control
>
Master
.
This sets the PSG to operate in master mode.